ELISA common problems and countermeasures problem possible reason Solution No color The time of reagent incubation did not follow the instructions. Determine whether the time for biotinylated antibody, linked HRP reagent or streptavidin-hrp is appropriate. Mix reagents in different kits or batches. Recheck the label of the reagent to ensure that all components belong to the reagent kit being used. Do not mix reagents in different kits or batches. Missed enzyme Check the operation process, be careful not to miss hrp enzyme contaminates sodium azide Use freshly prepared reagents and ban sodium azide There is a problem with the standard (if there is a signal in the specimen hole) Check the preparation process of standard products according to the instructions Use 1 new standard A container is not cleaned, leaving inactivated enzyme material Use the inner container of the kit as much as possible, and the other must be clean and reliable Prepare / re-dissolve antibodies using serum-containing buffer Reconfirm the selected reagent . Missing addition of developer a or b Observe the liquid level in the hole after adding the developer Incorrect reagent preparation / use Redo the experiment; follow the instructions strictly, and read the label clearly before each preparation and use. Buffer contamination Prepare fresh buffer Weak color rendering Products beyond the expiration date may produce a weak signal. Check the expiration date of the product Wrong volume and time of reagent added Make sure that the volume of each reagent used is correct and the addition time is appropriate. Reagents and samples are not balanced before use Allow reagents and samples to equilibrate at room temperature for about 10 minutes before use Shortening the incubation time can weaken the signal of the experiment. Check the time of incubation. Incubate the microtiter plate in an environment with varying temperatures. Determine the incubation temperature and avoid temperature changes. Contaminated reagents used Check if the reagent is contaminated. Standards / specimen preparation methods are not standardized Check the preparation of the standard / specimen. The standard should be reconstituted and diluted according to the instructions. Specimens stored at low temperatures avoid repeated freezing and thawing, and the use of hemolytic specimens is strictly prohibited. The sample is preserved with nan3, which inhibits the enzyme reaction Irregular substrate preparation Check the preparation of the substrate, such as the correct volume and proper mixing. Improper detection time Whether to test within the specified time. The instrument settings are incorrect and the filters do not match. Whether the instrument is set correctly and whether the filter selection is consistent. High background (background) Washing operation is not standardized Inadequate plate washing is often seen when using manual plate washing. It is best to use a plate washer, or use a bottle wash. Each well should be completely filled with washing buffer and should be poured quickly. If a plate washer is used, it should be calibrated and set to a volume sufficient to fill each well. The inside of the board should not touch the device. Check if there is any residual washing liquid in each well or the volume of sample added per well is accurate. Add 30 seconds of soaking between plate washes. Incubation temperature and time are inappropriate in the experiment Determine whether the incubation temperature and time for each experimental step are appropriate Enzyme added too much Check the pipette adjustment amount before adding enzyme. Check the dilution, if necessary, do potency determination test. Incomplete closure Check the calculated amount of blocking solution; extend the blocking time. Interfering substances in specimens or standards Make appropriate comparisons The developer is exposed to light for a long time or is contaminated Developers a and b should be removed from the refrigerator 10 minutes before use The entire batch of samples is placed for too long, and the samples are contaminated Samples should be kept fresh or stored at low temperature to prevent contamination Buffer contamination Prepare fresh buffer Reusable tips, not washed or incompletely disinfected and used for adding enzymes or color reagents Tips for single use as much as possible Too many signals: All the boards become regular blue Insufficient washing / washing steps were missed-unbound peroxidase remained. It is best to use a washing machine to fully wash Check if there is any residual lotion in the hole or if the sample volume is accurate. Substrate solution mixed too early and turned blue The timing of substrate mixing should be controlled and used immediately Too many enzyme conjugates Check dilution and perform potency determination if necessary The sealing film or reagent container is reused, resulting in the residual hrp, which makes the tmb substrate produce a non-specific blue. Use a new sealing film and use different reagent containers for each step Contaminated metals or hrp in the buffer Prepare fresh buffer High cv value (cv: coefficient of variation), flower plate Inadequate operation or insufficient washing Wash the plates, add samples and develop colors according to the instructions. Washing the board is particularly important, as mentioned above Dry board appears, no sealing film is used, and the sealing film is reused Make sure that the microplate should be kept moist between every two steps. Use sealing film to seal, pay attention to using new sealing film every step. The package is uneven due to operational errors or poor board quality (uneven bonding). Do not add other proteins to the diluted pbs Check the volume and time of the coating and blocking solution and the method of reagent addition. Check the microplates used, use elisa plates (do not use tissue culture plates) The pipette is inaccurate and the tip is reused. Check and calibrate the pipette. The tip must be changed every time a sample is taken. Review the steps of adding specimens to ensure the accuracy of each sample addition, and ensure that the sucked liquid is sucked and discharged according to the set volume. Pay attention to check the suction tip during continuous sampling to ensure the volume of the added liquid. Insufficient centrifugation of the sample, clotting or residual cell components in the reaction well Specimens are fully centrifuged at 3000rpm for more than 6 minutes. The use of coagulated (hemolytic) specimens is strictly prohibited Buffer contamination Prepare fresh buffer Standard curve is available, but the difference between the two points is poor (low or flat curve) Insufficient enzyme conjugate Check dilution and perform potency determination if necessary Capture antibody does not bind well to the plate Check the microplates used, use elisa plates (do not use tissue culture plates) Do not add other proteins to the diluted pbs Insufficient detection antibody Check dilution and perform potency determination if necessary Insufficient color development on the board Extended substrate incubation experiment Use recommended brand substrate solutions Careless operation Review the elisa operation process to eliminate any unauthorized modification procedures. Incorrect calculation of standard curve dilution Check the calculation and prepare a new standard curve The standard curve is good, but there is no expected positive signal No corresponding cytokines in the specimen Use internal control Repeat the experiment and reconsider the corresponding parameters of the experiment Specimen matrix detection Dilute the specimen at least 1: 2, or perform serial dilution to observe its recovery The standard curve is very good, but the interpretation value of the specimen is very high The level of cytokines contained in the specimen exceeds the detection range Dilute the specimen and experiment again When using hrp enzyme conjugate, tmb substrate and stop solution will turn green Insufficient color development of reagent in well Gently shake the board Edge effect Uneven working temperature Avoid incubating the microtiter plate in a changing temperature environment drift Discontinuity during the experiment The entire experiment should be operated continuously: prepare all standards and specimens properly before the experiment The reagent is not equilibrated to room temperature according to the instructions Before all reagents are added to the wells, make sure they are equilibrated to room temperature, unless otherwise stated in the instructions. Is it possible to change the experimental operation provided by the kit Steps? In order to ensure the highest sensitivity and specificity, general manufacturers have optimized the kits. To ensure the standardization of the experiments of each kit, they should follow the instructions. Can the reagents in different kits be mixed? Not allowed. Most reagents are specific in each batch of kits, if there is a problem, please contact the manufacturer or agent. Whether the volume of the specimen can be increased or decreased. The volume of the specimens to be added in the commercial kit is optimized, and it should be operated according to the instructions. It is not recommended to change the volume of the specimens added. Can I redefine the point of my own standard curve? can. There are recommended dilutions of the standard when preparing the standard curve in the instruction manual, which can change the dilution factor and increase the point of the curve, but it must be within the detection range. Invalid. Enamel Board,Combination Board,Magnetic Whiteboard,Melamine Whiteboard Dongguan Aoxing Audio Visual Equipment CO.,Ltd , https://www.aoxing-projectorscreen.com